Programmable Voltage Range |
0.10 kV-5.00 kV (10V/Step Resolution) |
|
Frequency Independent of Supply |
50 or 60 Hz |
|
Voltage Display Range and Accuracy |
0.10 kV-5.00kV ±1% ± 5 Counts |
|
Voltage Display Resolution |
0.01kV |
|
Current Display Range and Accuracy |
0.01mA-20.00mA ±1% ± 5 Counts |
|
Current Display Resolution |
0.01mA |
|
Selectable Range of Pass/Fail Levels |
0.01mA-20.00mA |
|
Maximum Current Output |
20.00mA @5kV |
|
Maximum Current Output Optional Arc Detection |
9 Levels |
|
Programmable Voltage Range |
0.10 kV-6.00 kV (10V/Step Resolution) |
|
Programmable Voltage Range Voltage Display Range and Accuracy |
0.10 kV-6.00kV ±1% ± 5 Counts |
|
Voltage Display Resolution |
0.01kV |
|
Current Display Range and Accuracy |
0.01mA-10.00mA ±1% ± 5 Counts |
|
Current Display Resolution |
0.01mA |
|
Selectable Range of Pass/Fail Levels |
0.01mA-10.00mA |
|
Maximum Current Output |
10.00mA |
|
Optional Arc Detection |
9 Levels |
|
DC Output Voltage |
250V, 500, or 1000V Selectable |
|
Display Range |
0.01MΩ-500MΩ |
|
Display Range / Accuracy |
0.03MΩ - 350 MΩ ± 5% ± 5 Counts |
|
|
350MΩ -500 MΩ Indication Only |
|
Display Resolution |
0.01MΩ |
|
Pass/Fail Level |
0.00MΩ-500.0MΩ |
|
|
|
|
Test Voltage |
Nominal 6V AC |
|
Frequency |
50 or 60 Hz (independent of supply) |
|
Display Range and Accuracy |
0-1500mΩ ±2% ±5 Counts |
|
Display Resolution |
1 mΩ |
|
Compliance Test Ranges |
|
|
Current/Load Resistance/Time |
5 A - 1000mΩ - Continuous |
|
|
10 A - 500mΩ - Continuous |
|
|
25 A - 200 mΩ - Continuous |
|
|
30 A - 150 mΩ - 60 secs |
|
|
40A - 100 mΩ - 60 secs |
|
Settable Ouput Current Range |
0.1A - 40.0A |
|
Selectable Range of Pass/Fail Levels |
0-1500 mΩ |
|
Test Voltage |
110V - 230V AC 20A nominal (Vin-Vout) |
|
Maximum Power Output |
Up to 5.0kVA (dependant on mains supply rating) |
|
Single Phase Power Measurement |
|
|
Display Range and Accuracy |
0.02kVA-5.00kVA ±2% ±0.02kVA |
|
Selectable Range of Pass/Fail Levels |
0.001kVA-5.000kVA |
|
Display Range and Accuracy Ratio |
0.000 -1.000 ± 0.030 |
|
Selectable Range of Pass/Fail Levels |
0.000-1.000 |
|
Display Range and Accuracy |
0.10mA-20.00 mA ± 1% ±5 Counts |
|
Display Resolution |
0.01mA |
|
Selectable Range of Pass/Fail Levels |
0.01mA-20mA |
|
Display Range and Accuracy |
0.02mA - 5.00 mA ± 1% ±5 Counts |
|
Display Resolution |
0.01 |
|
Selectable Range of Pass/Fail Levels |
0.02mA-5.00mA |
|
Measuring Device |
IEC/EN 60990 Fig 4 |
|
|
(IEC/EN 60990 Fig 3 or Fig 5 Upon Request) |