AC Voltage |
Selectable 115 or 230 VAC RMS |
|
Frequency |
50/60 Hz |
|
Size (HxWxL) |
300mm-200mm-370mm |
|
Weight |
15kg Approx |
|
Test Voltage |
Nominal 6V AC |
|
Frequency |
50 or 60 Hz (Independent of supply) |
|
Display Range and Accuracy |
0-1500mΩ ±2% ±5 Counts |
|
Display Resolution |
1 mΩ |
|
Compliance Test Ranges: |
|
|
Current/Load Resistance/Time |
5 A-1000mΩ - Continuous |
|
|
10 A-500mΩ - Continuous |
|
|
25 A-200 mΩ - Continuous |
|
|
30 A-150 mΩ - 120 secs |
|
|
40A-100 mΩ - 120 secs |
|
Settable Output Current Range |
0.1A-40.0A |
|
Selectable Range of Pass/Fail Levels |
0-1500mΩ |
|
DC Output Voltage |
250V, 500V, or 1000V Selectable |
|
Display Range |
0MΩ-500MΩ |
|
Display Range / Accuracy |
0MΩ-350MΩ ± 5% ± 5 Counts |
|
|
350MΩ-1000MΩ Indication Only |
|
Display Resolution |
1MΩ |
|
Pass/Fail Level |
0MΩ-1000MΩ |
|
Programmable Voltage Range |
0.10kV-5.00kV (10V/Step Resolution) |
|
Voltage Display Range and Accuracy |
0.10kV-5.00kV ±1% ± 5 Counts |
|
Voltage Display Resolution |
0.01kV |
|
Current Display Range and Accuracy |
0.01mA-20.00mA ±1% ± 5 Counts |
|
Current Display Resolution |
0.01mA |
|
Selectable Range of Pass/Fail Levels |
0.01mA-20.00mA |
|
Maximum Current Output |
20.00mA |
|
Optional Arc Detection |
9 Levels |
|
Programmable Voltage Range |
0.10kV-6.00kV (10V/Step Resolution) |
|
Voltage Display Range and Accuracy |
0.10 kV-6.00kV ±1% ± 5 Counts |
|
Voltage Display Resolution |
0.01kV |
|
Current Display Range and Accuracy |
0μA-9999μA ±1% ± 5 Counts |
|
Current Display Resolution |
1μA |
|
Selectable Range of Pass/Fail Levels |
1μA-9999μA |
|
Maximum Current Output |
9999μA |
|
Optional Arc Detection |
9 Levels |