OZU-4 head |
0.005-0.150 PPM |
|
OZL-4 head |
0.008-0.500 PPM |
|
OZG-4 head |
0.01-10.00 PPM |
|
Accuracy |
|
|
OZU-4 |
+/-0.005 |
|
OZL-4 |
+/-10% |
|
OZG-4 |
+/-15% |
|
Resolution |
|
|
OZU-4 head |
0.001 ppm |
|
OZL-4 head |
0.001 ppm |
|
OZG-4 head |
0.01 ppm |
|
Response time |
|
|
OZU-4 head |
70 sec. |
|
OZG-4 and OZL-4 heads |
60 sec |
|
Sensor type |
GSS |
|
Alarm |
2 set points, user configurable |
|
Outputs |
12-24 V, 4-20 mA, RS485, relays |
|
Temperature range |
23-104 F, (-5-40 C) |
|
Humidity range |
5% to 95% |
|
Power Requirements |
24VDC, 500mA (range 22-26 VDC) |
|
Size |
5˝x9x3ľ. Enclosure: 9-in W x 5-in H x 3.7-in D (230 W x 140 H x 95 D mm) |
|
Weight |
1.2 lbs (1.1 kg) |
|
Supporting documents |
User manual and NIST Traceable Calibration Sheet included |
|
Warranty |
1 yr |
|
External signals |
4 |
|
External signal functions |
Low Alarm, High Alarm, Control & Diagnostics |
|
Inputs |
Standby toggle |
|
Removable / replaceable sensor head |
Yes |
|
Sampling Pump |
12VDC, long life, brushless, diaphragm pump |
|
Enclosure casing |
Fiber reinforced polycarbonate |