MPS450患者模拟仪美国福禄克FLUKE

订货号 TS11363品牌 美国福禄克FLUKE信号发生器
产品简介
准确度:±2%
技术参数共 14 项
High-Level Output:(available on the BP3 connector) 0.2 V/mV ± 5 % of the ECG-amplitude setting
Rate Accuracy:± 1 % of setting
Pacer-Pulse Width:0.1
Pacing Rate:75 BPM
Waveforms:(Power-on default: atrial waveform)
PVC Focus:Left focus
Pulse Wave:30
Square Wave:2
Triangle Wave:2
Sine Wave:0.5
R-Wave Rate:30
R-Wave Width:20 to 200 ms (10-ms steps)
Accuracy Delta:± 10 %
Baseline:500
详细介绍
MPS450患者模拟仪性能特点
MPS450患者模拟仪Overview
The MPS450 is Fluke Biomedicals next-generation, portable, multiparameter patient simulator for your comprehensive testing training needs. Cardiac Output Fetal/Maternal ECG models are also available. (See “Models.”)
SMPS450患者模拟仪tard Features

  • ECG simulation/12-lead independent outputs
  • 43 arrhythmia selections
  • Four invasive BP channels including Swan-Ganz
  • Respiration simulation
  • Temperature simulation
  • BP synonization with ECG
  • Large, bright 4-line by 20-acter super twist display
  • R-Wave detection test
  • 9 V battery operation
  • External battery tests (9 V)
  • RS-232 port for computer control
  • Compact size
  • Front-panel soft-key user interface
  • Universal ECG connectors
  • High-level ECG output
  • Expansion port

MPS450患者模拟仪Optional Features

  • Cardiac-output simulation
  • Fetal/maternal ECG, direct simulations with intrauterine-pressure waveform (IUP)
  • Remote controller HHC3 (Hheld Controller)

 

Specifications
Power
 
Two 9 V alkaline batteries (eight hours min continuous power).
 
Optional battery eliminator
Normal-Sinus-Rhythm Waveform
ECG Reference:
ECG amplitudes specified for Lead II (calibration), from the baseline to the peak of the R-Wave; other leads proportional
Normal Sinus Rhythm:
12-lead configuration with independent outputs referenced to right leg (RL); output to ten Universal ECG Jacks, color-coded to AHA IEC Stards
Low-Level Amplitude:
0.05 to 0.50 mV (0.05-mV steps); 0.5 to 5.5 mV (0.5-mV steps) (Power-on default: 1 mV)
Amplitude Accuracy:
± 2 % of setting Lead II
High-Level Output:
(available on the BP3 connector) 0.2 V/mV ± 5 % of the ECG-amplitude setting
ECG Rate:
30, 40, 45, 60, 80, 90, 100, 120, 140, 160, 180, 200, 220, 240, 260, 280, 300 BPM (Power-on default: 80 BPM)
Rate Accuracy:
± 1 % of setting
ECG-Waveform Selection:
Adult (80 ms) or pediatric (40 ms) QRS duration (Power-on default: adult QR/P-R interval of 0.16 seconds)
Superimposed Artifact:
50 60 Hz, muscle, baseline wer, respiration
ST-Segment Elevation/Depression:
Adult mode only, -0.8 mV to +0.8 mV, in 0.1-mV steps; additional steps, +0.05 mV -0.05 mV (Power-on default: elevation 0 mV)
Pacemaker Waveform
Pacer-Pulse Amplitude:
1, 2, 5, 10 mV, ± 10 % (Power-on default: 5 mV)
Pacer-Pulse Width:
0.1, 0.5, 1.0, 1.5, 2.0 ms, ± 5 % (Power-on default: 1.0 ms)
Pacing Rate:
75 BPM
Waveforms:
(Power-on default: atrial waveform)
 
- Atrial
 
- Asynonous 75 BPM
 
- Dem with frequent sinus beats
 
- Dem with occasional sinus beats
 
- AV sequential
 
- Noncapture (one time)
 
- Nonfunction
Arrhythmias
Baseline NSR:
80 BPM
PVC Focus:
Left focus, stard timing (except where specified)
Supraventricular Arrhythmias:
(Power-on default: atrial fibrillation, coarse)
 
- Atrial fibrillation, coarse
 
- Atrial fibrillation, fine
 
- Atrial flutter
 
- Sinus arrhythmia
 
- Missed beat (one time)
 
- Atrial tachycardia
 
- Paroxysmal atrial tachycardia (PAT)
 
- Nodal rhythm
 
- Supraventricular tachycardia
Premature Arrhythmias (all one-time events):
(Power-on default: premature atrial contraction)
 
- Premature atrial contraction (PAC)
 
- Premature nodal contraction (PNC)
 
- PVC1 left ventricular
 
- PVC1 left ventricular, R on T
 
- PVC2 right ventricular
 
- PVC2 right ventricular, early
 
- PVC2 right ventricular, R on T
 
- Multifocal PVCs
Ventricular Arrhythmias:
(Power-on default: PVCs 6/min)
 
- PVCs 6/min (Power-on default)
 
- PVCs 12/min
 
- PVCs 24/min
 
- Frequent multifocal PVCs
 
- Bigeminy
 
- Trigeminy
 
- Multiple PVCs (one-time run of 2PVCs)
 
- Multiple PVCs (one-time run of 5 PVCs)
 
- Multiple PVCs (one-time run of 11 PVCs)
 
- Ventricular tachycardia
 
- Ventricular fibrillation, coarse
 
- Ventricular fibrillation, fine
 
- Asystole
Conduction Defects:
(Power-on default: first-degree heart block)
 
- First-degree heart block
 
- Second-degree heart block
 
- Third-degree heart block
 
- Right-bundle-branch block
 
- Left-bundle-branch block
ECG-Performance Testing
Amplitude:
0.05 to 0.50 mV (0.05-mV steps); 0.5 to 5.5 mV (0.5-mV steps) (Power-on default: 1 mV)
Pulse Wave:
30, 60 BPM, with 60 ms pulse width
Square Wave:
2, 0.125 Hz (Power-on default: 2 Hz)
Triangle Wave:
2, 2.5 Hz
Sine Wave:
0.5, 5, 10, 40, 50, 60, 100 Hz
R-Wave-Detection Waveform:
Haver-Triangle
R-Wave Rate:
30, 60, 80, 120, 200, 250 BPM (Power-on default: 60 BPM)
R-Wave Width:
20 to 200 ms (10-ms steps); additional steps, 8, 10, 12 ms (Power-on default: 10 ms)
Rate Accuracy:
± 1 %
Amplitude Accuracy:
± 2 %, Lead II (exception, ± 5 % for R-Waves ≤ 20 ms)
Respiration
Rate:
0 (OFF), 15, 20, 30, 40, 60, 80, 100, 120 BrPM (Power-on default: 20 BrPM)
Impedance Variations (∆Ω):
0.2, 0.5, 1, or 3 Ω (Power-on default: ∆ 1 Ω )
Accuracy Delta:
± 10 %
Baseline:
500, 1000, 1500, 2000 Ω, Leads I, II, III (Power-on default: 1000 Ω)
Accuracy Baseline:
± 5 %
Respiration Lead:
LA or LL (Power-on default: LA)
Apnea Selection:
12, 22, or 32 s (one-time events), or continuous (Apnea ON = respiration OFF) (Power-on default: 12 s apnea)
Blood Pressure
Input/Output Impedance:
300 Ω ± 10 %
Exciter-Input Range:
2 to 16 V RMS
Exciter-Input-Frequency Range:
DC to 5000 Hz
Transducer Sensitivity:
5 or 40 µV/V/mmHg (Power-on default: 5 µV/V/mmHg)
Pressure Accuracy:
± (2 % of setting + 2 mmHg)
Static Levels, P1: -
10, 0, 80, 160, 240, 320, 400 mmHg (Power-on default: 0 mmHg)
Static Levels, P2: -
10, 0, 50, 100, 150, 200, 240 mmHg (Power-on default: 0 mmHg)
Static Levels, P3: -
5, 0, 20, 40, 60, 80, 100 mmHg (Power-on default: 0 mmHg)
Static Levels, P4: -
5, 0, 20, 40, 60, 80, 100 mmHg (Power-on default: 0 mmHg)
Dynamic Waveforms, P1:
 
- Arterial:
120/80
- Radial artery:
120/80
- Left ventricle:
120/00
- Right ventricle:
25/00
Dynamic Waveforms, P2:
 
- Arterial:
120/80
- Radial artery:
120/80
- Left ventricle:
120/00
- Right atrium (central venous or CVP):
15/10
- Right ventricle:
25/00
- Pulmonary artery:
25/10
- Pulmonary-artery wedge:
10/2
- Left atrium:
14/4
Dynamic Waveforms, P3:
 
- Arterial:
120/80
- Radial artery:
120/80
- Left ventricle:
120/00
- Right atrium (central venous or CVP):
15/10
- Right ventricle:
25/00
- Pulmonary artery:
25/10
- Pulmonary-artery wedge:
10/2
- Left atrium:
14/4
Dynamic Waveforms, P4 (Swan-Ganz Sequence):
Right atrium (CVP); right ventricle RV); pulmonary artery (PA); pulmonary-artery wedge (PAW)
Respiration Artifact:
BP delta change from 3 to 16 mmHg
BP Output:
Mini DIN 7-Pin
Temperature
Temperature:
0 °C (32 °F), 24 °C (75.2 °F), 37 °C (98.6 °F), 40 °C (104 °F) (Power-on default: 0 °C / 32 °F)
Accuracy:
± 0.1 °C
Compatibility:
Yellow Springs, Inc. (YSI) Series 400 700
Output:
mini DIN 7-pin
Cardiac Output (optional)
Catheter Type:
Baxter Edwards, 93a-131-7f
Calibration Coefficient:
0.542 (0 °C injectate), 0.595 (24 °C injectate)
Blood Temperature:
37 °C (98.6 °F) ± 2 %
Injectate Volume:
10 cc
Injectate Temperature:
0 °C or 24 °C ± 2 % value (Power-on default: 0 °C injectate)
Cardiac Output:
2.5, 5, 10 lpm ± 5 % (Power-on default: 2.5 lpm)
Faulty-Injection Curve:
(Waveform for simulation available)
Left-to-Right-Shunt Curve:
(Waveform for simulation available)
Calibrated Pulse:
1.5 ° for 1 second (37 ° * 35.5 °) (Waveform for simulation available)
Repeatability:
± 1 %
Fetal/Maternal ECG (optional)
Maternal Heart Rate (Fixed):
80 BPM
Fetal Heart Rate (Selectable):
60, 90, 120, 140, 150, 210, 240 BPM (Power-on default: 120 BPM)
Fetal Heart Rate (IUP):
140 BPM at beginning, then varying with pressure
Intrauterine-Pressure Waveforms:
- Uniform acceleration (140-175-140 BPM; rate change lagging IUP contraction by 30 s)
 
- Uniform deceleration (140-100-140 BPM; rate change lagging IUP contraction by 30 s) (Power-on default)
 
- Early deceleration (140-100-140 BPM; no IUP lag time)
 
- Late deceleration (140-100-140 BPM, starting at IUP peak)
Wave Duration:
90 seconds, bell-shaped pressure curve, from 0 to 90 mmHg returning to 0, ± 4 mmHg (max)
IUP Period:
2, 3, or 5 minutes; manual (Power-on default: manual)
Computer Setup
Port:
Bidirectional (Data Communications Equipment) RS232
Baud Rate:
9600
Parity:
None
Stop Bits:
1
Data Bits:
8
Battery Replacement
 
Warning for low-battery condition (batteries to be replaced at this time)
Temperature
Storage:
-25 to 50 ºC
Operation:
10 to 40 ºC
Huity
 
80 % max relative huity
Dimensions
 
7.5” L x 7.25” W x 2” H
 
(19 cm L x 18.415 cm W x 5 cm H)
Weight
 
1.35 lb
 
(.612 kg)
……(完整参数见 PC 站)
同类产品
1000手持式起搏器分析仪
1000手持式起搏器分析仪
TS11450
咨询
MPS450患者模拟仪
MPS450患者模拟仪
TS11363
咨询
397任意波形发生器
397任意波形发生器
TS11533
咨询
电视信号发生器PM5410
电视信号发生器PM5410
TS02719
咨询
medSim300B患者模拟仪
medSim300B患者模拟仪
TS11362
咨询
PS420多参数患者模拟仪
PS420多参数患者模拟仪
TS11365
咨询
BP Pump2无创血压监护仪测试仪
BP Pump2无创血压监护仪测试仪
TS11413
咨询
CuffLink无创血压监护仪测试仪
CuffLink无创血压监护仪测试仪
TS11414
咨询
该型号由 北京中仪友信科技有限公司 供货 · 所属 试验部
010-51299017
工作日 8:30-17:30 · 节假日 13020017588
立即拨号复制型号
常见问题选购与报价
MPS450患者模拟仪 怎么报价?
该型号未公开标价,价格按采购数量与货期确定。致电 400-178-9117,报订货号 TS11363 即可取报价单。
美国福禄克FLUKE MPS450患者模拟仪有现货吗,货期多久?
货期随批次变动,下单前由工程师确认在册数量与到货时间。工作日 8:30-17:30 拨 010-51299017,非工作时间拨 13020017588。
怎么判断这个型号是否合用?
把测量对象、量程、精度要求和现场环境告诉工程师,可对照信号发生器类本页参数核对,必要时推荐同类在售型号。也可先索取实物照片与说明书。
开票与质保怎么算?
发票种类、质保期限与售后条款以购销合同为准,下单前与销售确认;后续可提供使用指导与返厂校准安排。
提示:本页参数由 就是要仪器网 整理,下单前请与工程师确认最新规格与货期。